Application Notes
Close
Fitting Inner Diameter (ID) Inspection
If
a view directly at the wall is desired, and the scope is a close fit
in the bore, the focus and illumination must be optimized for this
condition. A direction of view between 110° and 70° can be
used depending on the condition of the wall. The reason for choosing
a direction will depend on what is trying to be seen, and the
reflectivity of the wall.
Highly polished walls are very
difficult to view. The polished surface reflects the light away so
that almost no light gets reflected back into the scope. When looking
at welds that have a flatter finish, the scope will be able to
inspect the welded area, but the polished area will be difficult to
inspect.
A
straight ahead, or 0° direction of view can also be used to view
the walls of a close fitting bore. Sometimes a very wide field of
view is used to optimize the view of the wall and minimize the view
down the center of the bore. The 0° direction can be used to view
the entire ID at once without needing to scan the wall as in the side
viewing scopes described above.
Intersecting
Passages
If
coaxial lighting is needed to view into a very small passage, a
mirror tube may be needed. However, if the passage is large enough it
can be illuminated using separate prisms. If intersecting passages
are to be viewed, a mirror tube may be the best way to achieve the
direction of view. A mirror tube is a close fitting tube that slips
over the OD of the scope to position a mirror at the tip of the scope
to reflect the image and illumination at 90°. While mirror tubes
are very susceptible to dirt, they offer a coaxial lighting effect
that helps to view deeply into a passage at right angles to the scope
axis. When the passage is at an angle other than 90°, a mirror
tube may not be the appropriate solution.
A
flexible scope may be needed to enter the intersecting passage, and a
tool of some kind can be used to help the scope around the bend.
Normal practice has been to use articulating scopes for this, but
they are not necessary. Standard Zibra Corp products will do a good
job if they can be guided into the passages. They can also be
customized for special requirements. Please contact Zibra Corp to
discuss possible solutions.
Large
Cavities with Multiple Object Distances
A
large cavity usually requires a large amount of light. This simply
requires the largest possible scope. If the scope can be rigid it
will be a simple matter to focus at the required distances.
If
the scope needs to be smaller due to access limitations, multiple
access points should be investigated. Tools can be made to deliver a
smaller scope to defined points within the cavity. Auxiliary lighting
probes can be produced to help deliver light in addition to the light
from the scope probe.
If
a semi-rigid scope is desired for durability, a tip focusing scope
can focus at multiple distances. Another advantage of a tip focus
scope is the brightness can be better than an equivalent rigid scope.
If the access is restricted only in one dimension, probes can be
built to take advantage of the space available to maximize
illumination. In the case of a slit access, the probe can be built
with tubes placed adjacent to carry more lighting fibers and higher
resolution image optics.
Orbital
Welds in Piping
Welded
sections of sanitary piping or other high purity piping requires a
superflex scope that can make the sharp bends, look at the wall and
rotate 360° to view entire welds. A straight ahead view with a
wide field can be used, but the detail is not as good, and some of
the weld can be shadowed depending on the shape of the weld bead. The
multiscope® superflex scope has a range of centering tools that
can be added to the tip to maintain alignment in the pipe and focus
distance to the weld. These can inspect 1/4ӯ pipe to 4ӯ
pipe.
Standard
diameters are 2mm, 4mm, 6mm and 9mm. The 9mm diameter is built with
sheathing that is stiffer than the 2mm, 4mm and 6mm diameters. This
allows it to be pushed into larger piping.
Very
Close Focus
Close
focus requirements can be met with customized lenses. Our in-house
lens design capability means that we can accommodate your
requirements with a minimal lead time.
Fiber-based
scopes are very sensitive to the focus of the tip lenses. The focus
of the image at the tip of the scope cannot be changed once the scope
is built, and it is not changed by refocusing the video or eye
lenses. Rigid scopes use a more traditional relay system where the
image from the tip lenses is passed through a series of focal points
along the length of the probe, so a refocus of the rear lenses will
change the working distance at the tip of the scope.
In
a very close focus (0-3mm) application, the depth of focus is usually
reduced so that controlling position of the probe becomes more
important. Illumination can sometimes be difficult because a certain
amount of space is required for the lighting to cross into the
inspection area. Mirror tubes are sometimes used in close focus side
viewing applications because the airspace between the tip of the
scope, the mirror and the object allows the illumination to converge
at the target area.
Zibra
product part numbers for probes include the focus distance. We always
take focus requirements into account when building scopes. See our
page covering the part
number descriptions
for information on how to specify your focus requirement. Do not
hesitate to e-mail or call if you have either special requirements or
need to discuss the options for your particular needs. The close
focus scopes do require extra thought in specification and usage.
Infrared
Applications (IR)
Most
cameras have filters that eliminate the IR for color balance reasons.
When this filter is removed IR sensitivity is increased with some
corresponding loss of color balance.
Most
light sources generate plenty of IR, but it may be filtered out to
prevent burning of fibers. Filter(s) may need to be removed.
Illumination fiber in the borescope will be heat damaged if too much
IR is present, or cooling of lightguide tip is not adequate.
Scopes
built with fused quartz image fiber may lose image quality in the IR
spectrum since the fiber is optimized for use in the visible light
spectrum. Leached image bundles are recommended for IR use, or
verification that fused quartz will perform well as applied is
advised.
Fused
quartz fiber being used in radioactive environments will degrade more
slowly in the Infrared region.
Ultraviolet
(UV) Illumination Applications
Typically
this fiber has a lower acceptance angle and requires a narrow field
of view in the image optics to match the illumination pattern. UV
probes can be built for MilliscopeII® and Multiscope® with
continuous lightguide specification(CL). This construction has a
continuous illumination fiber bundle from the lightsource to the
probe tip, maximizing UV throughput. If a detachable illumination
bundle is desired, a Lumatech liquid lightguide can be used from the
lightsource to the probe. Lumatech also produces a high quality UV
lightsource.
Output
measurements have been made on the following scopes:
4mmØx90°
direction rigid probe ~4000 µW/cm² at 35mm.
1.7mmØ
x90° direction rigid probe ~300µW/cm² at 10mm.
These
tests were done using the Lumatech UV source and should serve as a
guideline only.
High Magnification
Probes
can be specified to create a microscope effect in the borescope.
These can be used to inspect fine features in areas where a
microscope cannot be used.
Fiberoptic
connectors are an ideal application for this. The fiber surface can
be inspected for dirt or polish flaws. The depth of focus is very
limited, and some type of locating fixture is required.
Magnifications of 100X-200X can be achieved. These can be built in
rigid, flex or semi-rigid.
Sheathing
Types
Rigid
and semi-rigid scopes are almost always sheathed using 304 stainless
steel tubing.
Flexible
scopes have a wide range of sheathing options. Plastic sheathings
include nylon 6, nylon 11, Teflon and polyimide. Reinforced
sheathings can have nylon braid, stainless braid and tungsten braid
on the outside or monocoil for crush resistance on the inside.
Illumination
bundles are normally sheathed in squarelock stainless tubing with
nylon or PVC extruded on the OD. MilliscopeII
illumination bundles are normally sheathed in a heavy wall nylon
tube.
High
Temperature
We
have experience building high temperature scopes for applications
that require insertion through difficult paths and static use.
Continuous bending shortens life, and sometimes causes unpredictable
failures. All high temperature scopes work better the less they are
cycled, and all will have some finite life. This varies by
application, and can sometimes vary unpredictably.
The
fiber bundles have chemical treatments applied that enhance
flexibility, and depending on the manufacturer, these will bake off
over time. If the scope is static, it is not usually a problem. These
can be built up to approx 3M in length.
Rigid
high temperature scopes have more choices. Cooling is easier to
apply, and they can be built using fiberoptics without regard to
bending issues. They can be built with conventional relay lens system
for ultimate resolution, but if un-cooled, they must be built without
any optical cement joints if working above ~400F. This will make the
images more likely to be flawed by incoming debris from the operating
environment.
Shaped
/ Shape Memory Probes
The
shape is put into the tube using a heating process. The scope can be
straightened with an outer sheath, and then deployed by retracting
the sheath, or it can be used without the sheath. This type of scope
is not advisable for general use where the shape memory properties
are not needed because the tubing is more expensive and has a thicker
wall which limits lighting possibilities. A scope of this type can
replace a much more expensive and fragile articulating scope.
Conventional
semi-rigid scopes can also be permanently bent into shapes specified
by the customer. These scopes will remain bent and will not
straighten or deflect like shape memory probes will.
Medical
Applications
Zibra
Corp builds medical instruments per customer specifications with
rapid design to delivery times. We are an FDA registered manufacturer
who can produce autoclavable micro endoscopes as well as other
medical optical devices. As with our industrial products, we can
offer optical design, mechanical design, prototyping, low and high
volume production all under one roof. Our recent agreement with
Biovision Technologies (www.biovisiontech.com)
allows us to concentrate on design and prototyping medical products.
Biovision can produce them in volume using their FDA and ISO
certified facilities.
Video
Equipment / Lightsources
Zibra
Corp sells complete video systems to support our borescope customers.
We recommend Imperx
frame capture cards for use with laptop computers and Elmo cameras
for multi- function capabilities. We recommend Elmo and TVS monitors.
We can quote items from any of these manufacturers.
We
have adapters for common digital cameras that allow for capture, zoom
and display on standard monitors.
We
have NTSC to VGA converters that allow camera output to be displayed
on a computer monitor (computer disconnected). These are commercial
items and can be also be found from other suppliers.
The
Welch-Allyn 24W SOLARC® lighsource which is ideal for the small
fiber bundles found in our scopes. This comes standard with an ACMI
port. We add our own attenuator design to improve the low light
cutoff through the scope. The lamp is a proprietary Welch-Allyn
product and the power supply is an external transformer which is
120/240 AC input rated and 12v DC,4A output.
We
also sell the Schott 150W ACE-1 halogen source for more general use.
This comes with an aluminum adapter blank that can be modified to fit
any size fiber bundle or adapter. The lamp is an EJA 12v, 150W and
the power input is 120VAC.